         | CATS Test Results
Here is a description of the most common test results that you may find on a CATS Summary Printout:
PASS This indicates that all parameters in the test passed their specified limits. Notice that the specified limits may have been changed from the default specifications.
FAIL One or more parameters did not pass the specified limits.
WARNING All parameters passed the specified limits, but one or more parameters got very close to the limit. The definition of what is considered to be close to the limit is specified individually by each manufacturer. This feature is not available until software version 2.20.
No Limits This indicates that the test was done without limits. This means that all parameters will be indicated as grey no matter what they measured.
ABORTED This indicates that the test was aborted before it was finished. We cannot say anything about the test result since the test was not completed.
FAIL (ABORTED) The test was aborted before it was finished, but one or more parameters were outside the specified limits before it was aborted.
Measurement Error Something went wrong during the test (maybe the player lost track). We cannot say anything about the test result since the test was not completed.
FAIL (Measurement Error) Something went wrong during the test. Maybe the player lost track, but one or more parameters were outside the specified limits before it failed.
Calibration Time Overdue This means that one or more modules in the test have not been calibrated within the time frame set up by the manufacturer. Notice that it cannot be guaranteed that the modules are calibrated just because this text is not there - it is possible to disable this check.
AudioDev AB: Kabingatan 9, SE-212 39 Malmö, Sweden, Tel: +46 40 690 49 00, Fax: +46 40 690 49 90, info@audiodev.com Partner Network Login
|
 |
|
| |
|